I just changed an SDG board in my system, now all my cal values are floating.  What should I do?

If SDG that was changed was in the last position connected, check for termination resistors RM36, RM8 & RM6;  They should be installed on the new board and can be taken from the board which was removed.  If the board is not the last spindle, confirm there are no other resistors in these locations in the remaining spindles.

 

 

 